The cryosauna installation shall include an oxygen detection system set to activate an alarm at 19.5% O2 by volume or sooner. The detector should be installed near the sauna cabin, in proximity to the occupant’s breathing zone. The alarm shall provide both audio and visual notification to notify both the occupant and operator when oxygen levels fall below the alarm set point.
Effects & Symptoms
• Inert cryogenic liquids like nitrogen, when vaporized, produce a large volume of cold, inert gas, displacing air, causing oxygen deficiency, and may accumulate in low points.
• Storage and handling of Nitrogen must meet the applicable requirements of NFPA 55, Compressed Gases, and Cryogenic Fluids Code 2013.
• Customers must consult with local and state agencies on all building, installation, and usage permitting and licensing requirements.
Cryosauna & Cryogenics
Suggested Gas Detectors, Control Panels & Accessories
Fixed gas detection for oxygen depletion, shut-down control and notifications.